Curve Tracer
371A
Specs
Characteristics
Collector
Supply
Modes
-
High Current: 250 µs pulses with max. peak of 30 volts.
High Voltage: full rectified sine with max. peak of 3000 volts
(positive and negative polarities for both modes).
|
Range |
Peak
Current |
Peak
Power |
High
Current |
30 V |
400
A |
3
kW |
(pulsed) |
30 V |
40
A |
300
W |
High Voltage |
3
kV |
40 mA
±20% |
30
W |
|
3
kV |
4 mA
±20% |
3
W |
|
300 V |
4 mA
±20% |
300
mW |
|
300 V |
0.4 mA
±20% |
30
mW |
Measurement
Characteristics
Collector
Current
Maximum
Power Setting |
Vertical
Range (in 1-2-5 increments) |
3
kW |
1
A/div to 50 A/div |
300
W |
500 mA/div
to 5 A/div |
30
W |
100
µA/div to 5 mA/div |
3
W, 300 mW |
10
µA/div to 500 µA/div |
30
mW |
1
µA/div to 50 µA/div |
Test
Fixture - Designed to allow easy connection to a variety of devices.
It has a safety enclosure to assure operator protection. Special patch
cords are provided for connecting large devices.
Step Generator
Modes
-
Current (stair step): DC, 500 µs ±10% of pulse.
Voltage (stair step): DC.
Step
Range -
Current: 1 mA/step to 2 A/step pulsed (high current mode). 1 µA/step
to 2 mA/step DC (high voltage mode).
Voltage: 200 mV/step to 5 V/step.
Offset: ±10X step amplitude (5X in 5 V/step and
2 V/step settings).
Safety
-
UL1244, CSA231, HD401 S1 |